PRODUCT APPLICATION
Tests on two temperature zones for different temperature shocks and changes within the performance of the equipment!
PS: Switching between high and low temperature zones is carried out by the cylinder, and there is a hanging basket inside the sample area
Use outside of the specified purposes may result in personal injury and damage to equipment!
PRODUCT PARAMETERS
product name | Thermal shock test chamber (two-slot type) |
model | CTS-100L |
Sample Restrictions | A) Testing and storage of samples of flammable, explosive and volatile substances B) Testing and storage of samples of corrosive substances Testing or storage of biological samples C) Testing and storage of samples of strong electromagnetic radiation sources |
Effective volume | About 100L |
Inner box size | W(width)500 mm*H(height)400 mm*D(depth)500 mm |
Test box outer box size | W(width)1500mm*H(height)18000mm*D(depth)1900mm, height includes raised height Tips: The external dimensions shall refer to the actual product! |
Conditions | No load (partial parameters as per notes) No sample |
Test chamber shock temperature range | Low temperature range -40 to -10℃ (or negotiated, -55℃/-65℃ optional), high temperature range +60~+150℃ |
High temperature box preheating temperature range: | +RT℃至+180℃ |
Low temperature box precooling temperature range: | -70℃至-10℃ |
Temperature fluctuation | ≤±0.5℃ |
Temperature uniformity | ≤±2℃ |
Heating time of high temperature box | +RT℃ rises to 150℃ for about 40 minutes, the average nonlinearity of the whole process |
Low temperature box cooling time | +RT℃ drops to -70℃ for about 70 minutes, and the average nonlinearity of the whole process |
Temperature Overshoot | ≤±2℃ |
Maximum load | 3KG plastic IC does not generate heat, does not include a test stand! |
Impact transition time | Stable within 5 minutes (temperature conversion, i.e. hanging basket switching time 5S) |
Minimum impact time of high and low temperature sections | 30 minutes |
Noise | ≤75dB (A sound level) |
Satisfaction test standards |
1. GB/T 2423.1-2001 Test A: Low temperature test method; 2. GB/T 2423.2-2001 Test B: High temperature test method; 3. GB/T2423.22-2002 Test N: Temperature change test method 4. GJB/150.3-1986 High temperature test 5. GJB/150.4-1986 Low temperature test 6. GJB/150.5-1986 Temperature shock test |
Outer Box | Cold rolled steel plate electrostatic spraying |
Inner box | SUS304# stainless steel plate |
Insulation layer | Rigid polyurethane foam and glass wool |
Door | Test box single door type |
Sealing | Original imported silicone sealing strip |
Switching valve | A. It is divided into high temperature tank, low temperature tank and test tank. The high temperature tank is used for energy storage, and the test tank is the user's experimental area. B. The high temperature tank and the test tank are separated by a switchable valve. The corresponding valve is automatically opened when impacted. C. There is also an independent ventilation valve, which operates to introduce ambient air when exhausting (restoring to normal temperature) |